Analytical HPLC 37-AHP100

HPLC Analytical system is equipped with a flow rate range of 0.0001 to 10 ml per minute. Integrated with a single and dual wavelength UV-Vis detector with a range of 190 to 800 nm that analyzes sample’s absorption at different wavelengths and a high pressure pump providing continuous consistent flow. Built in column heater provides reproducible retention time.

Pump Specifications
Model No HPP-A1
Flow rate range 0.0001 ~ 10 ml / min
Maximum pressure 50 MPa , 0 ~ 2 ml / min
Column size 4.6 x 250 mm11
Sample size 5 µm
Pressure pulsation ≤ 1 % at 10 MPa
Flow rate accuracy ± 0.15 % at 1 ml / min
Flow rate repeatability % ≤ 0.1
Gradient accuracy % ± 1.5
Gradient repeatability % ≤ 0.2
Dimension 460 x 310 x 180 mm
Power supply 90 V
Weight 11.6 kg
Detector Specifications
Model No HPD-A20 HPD-A21
Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
Wavelength accuracy ± 0.5 nm ± 2 nm
Wavelength repeatability 0.15 nm 0.3 nm
Band width 8 nm 8 nm
Optical path 10 nm 10 nm
Deuterium lamp 190 ~ 400 nm 190 ~ 400 nm
Tungsten lamp 400 ~ 800 nm 400 ~ 800 nm
Baseline noise ( Static ) Au ≤ ± 0.35 x 10 -5 ≤ ± 0.5 x 10 -5
Baseline drift ( Static ) Au / h ≤ 1 x 10 -4 ≤ 1.5 x 10 -4
Minimum detection limit ≤ 4 x 10 -8 ≤ 4 x 10 -8
Power 100 W 100 W
Power supply 90 V / 50 ~ 60 Hz 90 V / 50 ~ 60 Hz
Dimension 460 x 315 x 180 mm 460 x 315 x 180 mm
Weight 9.5 kg 9.5 kg
Autosampler Specifications
Model No ASI-A10
Syringe volume 1µl , 2.5 µl , 10 µl , 100 µl , 250 µl, 500 µl
Tray capacity ( removable rack ) 176 vials , 2 ml
Injection Speed 1 ~ 500 µl / s
Injection mode Full loop / partial loop
Analysis time 0 ~ 999 min
Wait time 12 s
RSD ( partial loop ) ≤ 0.5 %
RSD ( full loop ) ≤ 0.25 % ( typically ≤ 0.15 % )
Linearity ≥ 0.9999
Carry over ≤ 0.05 % ( with standard wash )
Power 55 W
Dimension 310 x 700 x 690 mm
Weight 15 kg
Standard Accessories
Accessory Quantity
Pump 2
UV-Vis detector ( without flow cell ) 1
Analytical flow cell 1
Injection valve 1
20 µl sample loop 1
4.6 x 250 mm ( 5µm ) analytical column 1
Solvent tray 1
Binary gradient kit 1
Mixing chamber 1
Optional Accessories
Accessory Quantity
Autosampler 1
Fraction collector 1
Column oven 1
  • Compression compensation ensuring gradient repeatability
  • Low pressure gradient
  • Detector ( Deuterium or tungsten lamp )
  • Automatic sample injector ( Autosampler )
  • 10 point calibration pump for linear pressure control
  • Error alarm function for over pressure ,under pressure and leakage

Used in pharmaceutical analysis, food nutrient analysis of amino acids and sugars, detection of food contaminants, environmental pollutant and chemical product purity analysis