Anesthesia Machine LE-ANM101 Catalog

Download

Overview :

lab Expo Anesthesia Machine LE-ANM101 is designed to deliver dual gas support (O₂ and N₂O) with precise control. It offers a flow range of 0.1 to 10L/min for both gases, ensuring efficient pneumatic operation. Equipped with a 5.7-inch LCD display, integrating respiratory control functions and real-time monitoring parameters. Integrated flow meter system and gas management ensure reliable and stable performance. Our Anesthesia Machine is suitable for hospitals, clinics, and surgical centers.

Specifications:
Ventilation System

Microprocessor Controlled with 5.7-inch LCD Display

Modes

IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH

Tidal Volume

50 to 1500ml

Respiratory Rate

2 to 99 bpm

I:E Ratio

2;1 to 1:8

Safety Features

O₂/N₂O Interlock (min 25% O₂)

Automatic N₂O Cutoff (< 0.2 MPa O₂ pressure)

Comprehensive Alarm System (O₂ Concentration, Airway Pressure, Apnea)

Vapourizer

Auto-Compensating (Temperature/Pressure/Flow)

Vapourizer Options

Compatible with Halothane/ Enflurane/Isoflurane/ Sevoflurane

Vapourizer Concentration

Concentration Range: 0 to 5 vol%

Gas Delivery

Dual Gas Input (O₂ & N₂O: 0.27 to 0.55 MPa)

Gas Delivery Range

Flow Meters: 0.1 to 10 L/min (Both Gases)

Emergency Gas Delivery

Emergency O₂ Flush: 25 to 75 L/min

Oxygen Monitoring

Continuous O₂ Monitoring (21 to 100%)

Pressure Monitoring

Airway Pressure Monitoring (1.0 to 6.0 kPa)

Flow Monitoring

Real-Time Waveforms (Pressure/Flow)

Patient Compatibility

Adult/Pediatric Convertible

Bellows System for Both Patient Types

Multiple Circuit Modes (Closed/Semi-Closed/Open)

Power Supply

AC 220V ±10%, 50Hz ±1Hz

Features:

  • Microprocessor-controlled ventilator with 5.7-inch LCD display
  • Dual gas input O₂ & N₂O: 0.27 to 0.55 MPa pressure range
  • Flow meters: 0.1 to 10 L/min for both gases
  • Tidal volume adjustable from 50 to 1500 ml
  • I:E ratio adjustable from 2:1 to 1:8 ratio settings
  • Airway pressure monitoring: 1.0 to kPa measurement range

Applications:

FunctionImplementationSurgical suitesDual gas input (0.27 to 0.55 MPa) supports O₂ and N₂O deliveryICU ventilationTidal volume range 50 to 1500ml enables adult and pediatric careOperating roomsRespiratory rate 2 to 99 bpm controls breathing cyclesRecovery wardsEmergency O₂ flush 25 to 75 L/min provides immediate supportClinical monitoringContinuous O₂ monitoring (21 to 100%) ensures patient safety

www.labexpo.com