Biological Safety Cabinet Class II B2 56-BSC201 has safety level 2 enclosed cabinet unit where there is no recirculation inside the workspace; 100 % air is exhausted as they are hard connected to exhaust system. This is used to handle volatile toxic chemicals and radio nucleotides where chemical vapor re-circulation into the work zone is not permitted. It meets the requirements for the protection of product, personnel and the environment allowing for the safe handling of pathogens, contaminants or other potentially hazardous materials.
| Air exhaust | 100% |
| Glass door orientation | 10° tilted |
| Working area | 1000 x 650 x 620 mm |
| Overall dimension | 1195 x 720 x 1950 mm |
| HEPA filter | 0.3 micron |
| Air pressure | Vertical negative |
| Cleaning level | 100 level |
| Ascending wind speed | ≥ 0.55 m/s |
| Descending wind speed | ≥ 0.35 m/s |
| Noise | ≤ 58 dB |
| Half ventilation peak value | ≤ 3 µm |
| Personnel protection | CFU ≤ 10 |
| Product protection | CFU ≤ 5 |
| Cross contamination | CFU ≤ 2 |
| Fluorescent lamp | 20 W x 2 |
| UV lamp | 20 W x 1 |
| Power | AC 220 V / 50 Hz |
| Air supply filter dimension | 955 x 545 x 50 mm |
| Weight | 300 kg |
FunctionImplementationVolatile chemical proceduresFully exhausted airflow preventing vapor return to workspaceRadionuclide handlingHard-connected exhaust for trace radioactive material controlBiological hazard containmentCFU ≤10 personnel protection under validated airflow conditionsSterile sample processingCFU ≤5 product protection within controlled clean zone