Biological Safety Cabinet Class II B2 56-BSC202 Catalog

Download

Overview :

Biological Safety Cabinet Class II B2 56-BSC202 safety level 2 enclosed cabinet unit where there is no recirculation inside the workspace; 100 % air is exhausted as they are hard connected to exhaust system. This is used to handle volatile toxic chemicals and radio nucleotides where chemical vapor re-circulation into the work zone is not permitted. It meets the requirements for the protection of product, personnel and the environment allowing for the safe handling of pathogens, contaminants or other potentially hazardous materials.

Specifications:
Air exhaust 100%
Glass door orientation 10° tilted
HEPA filter 0.3 micron
Air pressure Vertical negative
Cleaning level 100 level
Ascending wind speed ≥ 0.55 m/s
Descending wind speed ≥ 0.35 m/s
Noise ≤ 58 dB
Half ventilation peak value ≤ 3 µm
Personnel protection CFU ≤ 10
Product protection CFU ≤ 5
Cross contamination CFU ≤ 2
Fluorescent lamp 30 W x 2
UV lamp 30 W x 1
Power AC 220 V / 50 Hz
Air supply filter dimension 1290 x 554 x 50 mm
Working area 1300 x 650 x 620 mm
Overall dimension 1495 x 720 x 1950 mm
Weight 340 kg

Features:

  • 100% HEPA-filtered air exhaust with no workspace recirculation
  • 0.3 µm HEPA filtration sustaining Level 100 cleanliness
  • Upward airflow ≥0.55 m/s with 0.35 m/s vertical downflow
  • Negative pressure chamber directly ducted to external exhaust
  • 1300 × 650 × 620 mm stainless steel work area with LCD controller

Applications:

FunctionImplementationToxic vapor proceduresFully exhausted airflow preventing chemical vapor returnRadioisotope applicationsHard-ducted system for radionuclide containmentBiological risk containmentCFU ≤10 personnel protection in validated airflow fieldSterile culture operationsCFU ≤5 product protection within clean airflow zoneDiagnostic and QC environmentsMid-size workspace supporting regulated lab tasks