IR Carbon and Sulphur Analyzer is a unit for determination of sulphur and carbon equipped with two treatment units including a high frequency induction burner and flexible temperature settings with automatic control. With an analysis time of 25 to 60 seconds and oxygen purity of 99.5 %, the multi-shielded isolation circuit prevents the machine for high frequency interference. High speed 24-bit sampling improves sensitivity and accuracy.
Used to measure mass fraction of carbon and sulfur in steel, iron, casting core sand, cement, ore, catalyst, magnetic materials, ceramics, inorganic matter, graphite, refractory materials, battery materials, plants and other materials.
| Carbon measurement | 0.0001 % to 10.0000 % |
| Sulphur measurement | 0.0001 % to 3.5000 % |
| Sampling per second | 28 times |
| Relative humidity | < 90 % |
| Oxygen purity | ≥ 99.5 % |
| Room temperature | 10 ~ 30 °C |
| Oxygen input pressure | 0.18 MPa at 5 % |
| Sensitivity | 0.1 ppm |
| Carbon detection pool | Can add low carbon pool |
| Sulphur detection pool | Can add high sulphur pool |
| Analysis time | 35 seconds ( can be set 25 to 60 seconds ) |
| Weighing precision | 0.0001 g |
| Power | ≥ 2.5 kV |
| Voltage | 220 V |
| Oscillation frequency | 20 MHz |
| Frequency | 18 Hz |