0

Cleveland Open Cup Flash Point Tester 52-OFP102

Cleveland Open Cup Flash Point Tester 52-OFP102 is equipped with English man-machine interface. It can preset expected flash point temperature and other parameters which is reasonably designed, safe and convenient to operate. It is designed and made as per ISO 2592, ASTM 92 “Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Methods).

Price: $7968

Ambient Temperature 10 ℃ to 40 °C
Relative Humidity ≤ 80 %
Temperature Measurement- Full Scale Ambient temperature to 300 °C
Temperature Measurement- Repeatability ≤ 8 °C
Temperature Measurement- Reproducibility ≤ 16 °C
Temperature Measurement- Resolution 0.1 °C
Test accuracy 0.002
Precision 0.005
Flame 3 to 4 mm
Dimensions 550 x 470 x 410 mm
Power Consumption ≤ 500 W
Power Supply AC 220 V ± 10 %, 50 Hz
  • Automatic program temperature controlling
    Automatic meteorological pressure standardization and altered value correction
    Electric flame ignition mode
    LCD display
    It can print out test data automatically Convenient and rapid

  • Cleveland Open Cup Flash Point Tester is used to determine flash point of petroleum products and bituminous materials, excepting fuel oil and materials having a flash point of lower than 79 °C

Open Cup Flash Point Tester