Request Quote
Scan the QR code below to connect with us on WhatsApp.
Request Quote
Scan the QR code below to connect with us on WhatsApp.
Gas Chromatograph 65-GC101 performance is enhanced by 8 orders temperature programming with precise temperature control across multiple areas. It includes column boxes, TCD, FID, and conversion furnaces. It has a temperature range of 5℃ to 420℃ at room temperature and accuracy better than ± 0.1 ℃. Our Chromatograph possess accurate pressure and flow measurements using pointer-type or electronic instruments.
| Operation | 192 × 64 periodic lattice liquid crystal |
| Temperature Control | 5℃ to 420℃ at room temperature (protection temperature 200℃) |
| Temperature Increment | 1℃ |
| Temperature Control Accuracy | <± 0.1℃ (measured at 200℃) |
| Temperature Control Area | 6 Ways |
| Temperature-Programmed Order | 8 Order |
| Programming Rate | 0.1 to 39℃/min (common type); 0.1 to 80℃/min (high speed) |
| Gas Control | Optional pressure flow control mode of the mechanical valve control mode, electronic |
| Gas Pressure Flow Measurement | Pointer type pressure meter or electronic pressure, flow pattern |
| Measuring Way | 16 Ways, Single module |
| Measuring Range | 0 to 0.3MPa (pressure); 0 to 500sccm (flow) |
| Measuring Resolution | 0.01kPa (pressure); 0.01sccm (flow) |
| Working Mode | EPC, EFC, Constant current mode, Constant voltage mode |
| Working Gas | Nitrogen, Air, Hydrogen, Helium, Argon |
| Detector Type | TCD and FID |
| TCD Sensitivity | S ≥ 2500mV•ml/mg (sixteen alkyl) (1, 2, 4, 8 times enlarged, optional) |
| TCD Baseline Noise | ≤ 20μV |
| TCD Baseline Drift | ≤ 30 μV/30min |
| TCD Linear Range | ≥10⁴ |
| FID Detection Limit | ≤ 1×10⁻¹¹g/s (sixteen alkyl) |
| FID Baseline Noise | ≤5×10⁻¹⁴A |
| FID Baseline Drift | ≤1×10⁻¹³A/30min |
| FID Linear Range | ≥10⁶ |
| Dimensions | 52 × 48 × 55 cm |
| Net Weight | 53 Kg |
FunctionImplementationMulti-component separation8-order temperature programming manages complex mixturesTrace compound detectionFID limit ≤1×10⁻¹¹ g/s enables low-level analysisGas composition analysisTCD sensitivity supports quantification of permanent gasesPharmaceutical testingPrecise temperature and flow control ensure reproducible resultsEnvironmental sample evaluationStable baseline drift and noise maintain measurement accuracy
Computer 1080 USD
27 samples of Headspace autosampler 18880 USD
Printer 840 USD