Request Quote
Scan the QR code below to connect with us on WhatsApp.
Request Quote
Scan the QR code below to connect with us on WhatsApp.
Lab Expo Nucleic Acid Extractor LE-NAE101 offers a processing volume range from 30 to 1000 μl for low and high-throughput sample. Features multi-modes and multi-gears mixing to enable efficient nucleic acid extraction. Built-in UV disinfection module to ensure a sterile working environment. Supports remote control via android devices and on-device operation through screen and buttons. Our Nucleic Acid Extractor comes with magnetic beads-based extraction kits for high-purity nucleic acid isolation.
FunctionImplementationClinical diagnostics labs30 to 1000 μl processing volume supports varied sample throughputQuality control workflows≤ 1% magnetic bead residue ensures high-purity extractionResearch facilitiesRT to 120℃ lysis and elution heating enables protocol flexibilityIndustrial testing environments600 W power consumption allows stable benchtop operationInstitutional research settingsUp to 15 sets of programs stored locally enables workflow continuity
| Processing Volume | 30 to 1000 μl |
| Sample Volume | 200 μl |
| Magnetic Bead Residue | ≤ 1% |
| Applicable Consumables | 96-Well Plates. 6- Tube Strips |
| Heating Temperature | Lysis Heating: RT to 120℃ Elution Heating: RT to 120℃ |
| Operating Temperature Range | 10 to 30℃ |
| Operating Humidity Range | 20% to 85% |
| Mixing Mode | Multi-Modes and Multi-Gears Mixing |
| Reagent Type | Magnetic Beads-Based Extraction Kits |
| Operation Mode | A. Remote Control via Smart Phone /Tablet (Android System) B. Screen/Button Operation |
| Experimental Storage | Up To 15 Sets of Programs Can Be Stored in The Instrument Android Applications Can Store > 500 Programs |
| Program Management | Programs can be created, edited, applied and deleted with high flexibility |
| Contamination Control | Built-in UV disinfection module |
| Power Failure Protection | In the event of an unexpected power failure, there are options available whether or not to continue the experiment |
| Connection Port Type | USB, Ethernet, USB Type-B |
| Network Connection | Mobile/Tablet Wifi Remote Control; Ethernet Remote Control Applicable |
| Power Consumption | 600 W |
| Power Supply | AC 100-240V, 50/60+1Hz; |
| Instrument Dimensions | 435 × 440 × 445 mm |
| Net Weight | 31.5 kg |