ICP spectrometer 60-ICP100

Lab Expo ICP spectrometer 60-ICP100 detects trace elements using argon plasma where the atomized liquid sample is injected. With an analyzing rate of 26 elements per minute results can be achieved with lower levels of detection and measurement of trace and ultra-trace elements. Elements can be traced from liquid and solid samples. Detection of liquid flow from the spray chamber by optical drain sensor offers exceptional sensitivity and accuracy.

$80,525.00 $ 67104

ICP spectrometer 60-ICP100 - Specifications

Wavelength range ( 3600 lines / mm ) 180 to 500 nm
Wavelength range ( 2400 lines / mm ) 180 to 800 nm
Temperature 20 ~ 28 °C
Elements per minute 26 elements
Humidity 70%
Focal length 1000 mm
Standard deviation RSD ≤ 1.0 %
Correlation coefficient ≥ 0.9998
Gas source Argon ( 99.99 % )
Plasma gas flow meter 100 ~ 1000 L / h
Auxillary gas flow meter 10 ~ 100 L / h
Carrier gas flow meter 10 ~ 100 L / h
Cooling water flow rate 5 liters per min
Cooling water temperature 21 ~ 26 °C
Water outlet diameter Ø 10 mm
Line dispersion rate 0.26 nm
Resolving power ≤ 0.008 nm
Incident slit 20 µm
Co-axial sprayer diameter 6 mm
Coil diameter 25 mm ( 3 turn )
Double room fog diameter 34 mm
Quartz tube torch diameter 20 mm
Mirror dimensions 78 x 105 x 16 mm

ICP spectrometer 60-ICP100 - Key Features

  • Wavelength range 180 to 500 nm with 3600 lines/mm grating
  • Elements per minute analysis rate of 26 elements
  • Temperature range maintained at 20 to 28 C operating conditions
  • Optical pathway Czerny-turner configuration for spectral dispersion
  • Low detection limit capability for trace element quantification

ICP spectrometer 60-ICP100 - Applications

FunctionImplementation
Trace element detection26 elements per minute enables rapid multi-element analysis
Liquid sample processingCo-axial sprayer diameter 6 mm ensures consistent nebulization
Wavelength selectionLine dispersion rate 0.26 nm supports precise spectral separation
Resolution capabilityResolving power ≤ 0.008 nm allows fine peak discrimination
Environmental stabilityTemperature range 20 ~ 28 °C maintains operational consistency

ICP spectrometer 60-ICP100 - Element Detection Limits

Element Be Mg Ca Sr Ba Sc Y Eu Yb
Wavelength ( nm ) 313.042 279.553 393.366 407.771 455.403 335.373 371.03 381.967 369.419
Detection limit < 1.0 < 1.0 < 1.0 < 1.0 < 1.0 < 1.0 < 1.0 < 1.0 < 1.0
Element Lu La Gd Dy Ho Er Tm Zn Cd
Wavelength ( nm ) 261.542 333.749 342.247 353.17 345.6 337.271 313.126 213.856 226.502
Detection limit < 1.0 < 3.0 < 3.0 < 3.0 < 3.0 < 3.0 < 3.0 < 3.0 < 3.0
Element Fe Ti Cu Mn V B Ni Cr Co
Wavelength ( nm ) 239.562 344.941 324.754 257.61 309.311 249.773 221.647 205.552 238.204
Detection limit < 3.0 < 3.0 < 3.0 < 3.0 < 3.0 < 5.0 < 5.0 < 5.0 < 5.0
Element Zr Mo Ag Au Si Al Ir P Hg
Wavelength ( nm ) 343.823 202.03 328.068 242.795 251.611 396.152 224.268 213.618 194.227
Detection limit < 5.0 < 5.0 < 5.0 < 5.0 < 5.0 < 10.0 < 10.0 < 10.0 < 10.0

ICP spectrometer 60-ICP100 - Catalog

Related ICP spectrometer

Lab Expo ICP Spectrometer units are designed for elemental analysis in environmental, chemical, and industrial testing. Models support ICP optical emission spectrometer applications with wavelength ranges from 180 to 900 nm and multi-element detection for icp oes spectroscopy and trace element analysis.